|
Business Accent
Identifying Benzodiazepines
Michael Zumwalt and John M Hughes narrate the
use of LC/MS in tracing Benzodiazepines.
The analysis of benzodiazepines is of great interest to forensic
and clinical toxicologists. Using the Agilent LC/MSD VL Quadrupole mass spectrometer
instrument in full scan mode, both spectral identification and quantitation
can be carried out simultaneously.
Benzodiazepines are an important class of drugs with a broad
range of therapeutic effects, including sedative-hypnotic, anxiolytic, muscle-relaxant,
and anticonvulsant. Because of their wide usage, benzodiazepines have the potential
for interaction with other central nervous system depressants which can result
in life-threatening or impaired driving situations. Benzodiazepines are now
among the most commonly-prescribed drugs, which increases their potential for
addiction and abuse, and often they are found in combination with other drugs
in drug-related fatalities or drug facilitated sexual assault cases. The analysis
of benzodiazepines is therefore of great interest to forensic and clinical toxicologists.
Benzodiazepines have been analysed using HPLC with UV detection,
gas chromatography with nitrogen phosphorus and electron capture detectors,
and gas chromatography/mass spectrometry (GC/MS). Many benzodiazepines are polar
and non-volatile, making them difficult, if not impossible, to analyse with
GC or GC/MS. Some of the compounds cannot be derivatised for improved chromatographic
behaviour. Furthermore, some of the newer benzodiazepines, like flunitrazepam,
have lowered therapeutic ranges and faster clearance, and therefore require
quantitation at lower levels. Liquid chromatography/quadrupole mass spectrometry
is ideally suited for these compounds because the technique does not require
derivatisation, thereby saving time, expense, and experimental difficulty. The
full-scan sensitivity of the Agilent liquid chromatograph/mass selective detector
(LC/MSD) allows for quantitation, identification and confirmation in a single
analysis.
Experimentally...
The LC/MS system consists of 1100 series vacuum degasser,
binary pump, autosampler, thermostatted column compartment, diode array detector
(DAD) with micro-flow cell, and LC/MSD quadrupole VL model. The DAD was used
primarily for method development. However, the UV detector in series with the
MS provides UV spectra which can also be used for identification when levels
are sufficiently high. Complete system control and data analysis was provided
by the Agilent LC/MS ChemStation.
Sample preparation
Samples were prepared using liquid-liquid extraction, which
is commonly used for these compounds for analysis by GC/MS. The only difference
from a GC/MS method is omitting the derivitisation step and reconstitution of
the final sample in the LC mobile phase, rather than in a volatile solvent for
GC injection. A 1 ml volume of blood, serum, or plasma, to which 100 µL
of internal standard solution (10 ng/µL) has been added, is added to 1
ml of saturated sodium borate solution, and the mixture is vortex-mixed. Ethyl
acetate (4 ml) is added and mixing is carried out on a rotary shaker for five
minutes, followed by centrifugation at 3400 rpm for five minutes. The upper
layer is transferred to a clean tube and evaporated to dryness. The residue
is reconstituted in 50 µl of the initial mobile phase and transferred
to an autosampler vial.
Results and discussion
|
A moderate amount of collision-induced
dissociation (CID) was used in this method by setting the fragment or
voltage in the method to a value 50V higher than the default value of
70V which minimises CID
|
The sample preparation used for many GC or GC/MS methods can
often be used for LC/MS just by omitting any derivatisation step and transferring
the final sample to LC mobile phase instead of using a volatile solvent. For
this experiment, Flumazenil, a benzodiazepine antagonist not found in samples
in this jurisdiction, was used as the internal standard due to the cost and
availability of deuterated analogues of some analytes.
The standard VL model of the LC/MSD is quite capable of carrying
out the analysis of benzodiazepines in blood. The SL model affords approximately
10x greater sensitivity if needed for other analyses, as well as multisignal
capability such as alternating positive/negative mode, SIM/scan mode, and low/high
fragmentation modes. The analysis was carried out in full scan acquisition mode
in order to quantitate the target analytes using extracted ion chromatograms
(EICs), and to alternatively confirm their identity using other ions in the
spectra as well.
A moderate amount of collision-induced dissociation (CID)
was used in this method by setting the fragment or voltage in the method to
a value 50V higher than the default value of 70V which minimises CID. This results
in spectra which contain more ions than just the pseudo-molecular ion. These
ions can then be used as confirming ions for EICs, as is the common practice
for EI GC/MS, and the spectra can be placed in a user created library for identification
of drugs using library search of API spectra.
The fragmentor voltage chosen is just high enough to produce
fragment ions by CID for confirmation, while attempting to preserve significant
signal for the intact molecule. For example, the m/z 268.1 in the spectrum of
flunitrazepam arises from fragmentation of the (M+H)+ ion at m/z 314.1. The
spectral behaviour is, of course, compound dependent. For instance, in the case
of temazepam, there is more signal for the sodium adduct at m/z 323.1 so that
ion is used for quantitation, while the fragment at m/z 255.0 is used for confirmation.
The protonated molecular ion m/z 301.0 can also be used as a confirmation ion,
as long as the protonated/sodiated ion ratios are constant for the analysis.
Compounds with oxygen-containing functional groups can show sodium adducts as
well as proton adducts; this complication can be avoided with the use of Atmospheric
Pressure Chemical Ionisation (APCI) in place of electrospray ionisation (ESI).
The limit of detection (LOD) at a S/N of 3:1 is approximately
10 ng/ml using this method for most of the target compounds and this model of
LC/MSD. The method as practised at the Montana State Toxicology laboratory uses
a 20 ng/ml (0.02 µg/ml) limit of quantitation (LOQ), a calibration range
extending to 1,000 ng/ml, and one or
more qualifier ions for each analyte.
With the specified sample preparation and instrument conditions
and a calibration range up to 1000 ng/mL, a quadratic treatment gives a better
curve fit than a linear treatment for most of these analytes (r2 >0.99).
Results of an actual case sample, in which alprazolam was found at a moderate
level of 128 ng/ml.
Conclusion
This experiment demonstrated the usefulness of LC/MS for
the analysis of benzodiazepines in blood. These compounds tend to be difficult
to analyse by GC-based techniques, but ionise well in API-electrospray, resulting
in excellent sensitivity, even in full scan mode using the lowest-cost model
of LC/MSD (VL). Blood is a difficult matrix to analyse, but the results here
show excellent quantitation and simultaneous identification using only 1 ml
of sample and a simple liquid- liquid extraction procedure used for GC/MS, without
the derivatisation. The CID spectra show strong basepeak signals used for quantitation
over three orders of magnitude, and CID fragment ions for ion ratio confirmation
and/or library search. The method could be used in SIM and with the more sensitive
LC/MSD SL model for any of the newer benzodiazepines which are found at lower
levels in blood.
A rapid, validated method has also been published for a large
number of benzodiazepines and related substances using the Agilent LC/MSD quadrupole
system. The method uses APCI rather than ESI, a liquid/liquid extraction procedure
similar to this one, CID with greater fragmentation, and several deuterated
internal standards. The publication describes the use of CID spectra and library
search for identification, and includes spectra of all the analytes under both
low and high fragmentation conditions.
(The authors are Senior Application Scientist and Senior
Applications Consultant, Mass Spectrometry of Agilent)
|